Performance parameters of GC 2000

GC 2000 is a special-purpose detection and analysis instrument for laboratories based on the basic principle of gas chromatography. It adopts the advanced electronic flow pressure control system, microfluidic plate control technology. high-precision independent temperature control system, and high-sensitivity detector, which, in combination with the flexible and friendly user interface, high-speed sampling frequency and signal processing speed, and networked data and control platform, meet the user’s requirements for the capability. reliability, stability and advanced nature of instrument analysis.

Chromatographic performance:

  • Retention time repeatability < 0.008%
  • Peak area repeatability <0.5%

Our Core Strengths

  • Supporting simultaneous installation:
    a.Two injection ports
    b.Four detectors
    c.Three detector signals
  • Electronic digitized data output of the advanced detector, enabling quantitative analysis of peaks over the full range of the detector in a single injection
  • Electronic flow control for all injection ports, detectors, and auxiliary gas flow circuits
  • Possible for up to 6 electronic flow control modules to be installed, to provide control over up to 18 gas circuits. Accuracy of pressure setpoint and control approaching 0.001 psi, providing more accurate retention time for low pressure analyses.
  • Electronic flow controller for capillary columns with five column flow control modes: Constant pressure, pulse, and programmed pressure rise and flow increase to the fourth order; possible to calculates the average linear velocity of the column.
  • Standardized pressure and temperature compensation to keep test results remain unchanged even when the laboratory
    environment changes.
  • Possible for users to automate leak checks before each run, or run the check function at any time during maintenance or diagnostics. Users are alerted when a leak is detected so that immediate action can be taken to correct and reduce system
    downtime.

Color touch screen interface :

  • 8-inch high-resolution full-color capacitive touch screen, with the resolution at 1280 (RGB) *720, color depth at 16.7M, and pixel pitch (mm) at 0.138*0.138.
  • Customized host control software based on the Android system, with the graphical interface, full touch for operation, the UI also graphical, in combination with intelligent functions such as self- diagnosis reminder, self-leakage detection, etc., reducing the difficulty of use and maintenance, and for easily grasping of the status of the instrument.

Column oven :

  • Specification: 28x31x16 cm, possible to hold up to 3 capillary columns.
  • Temperature range applicable to all columns and their separation requirements. Room temperature+4°C to 450°C Temperature setting accuracy: ≤0.1°C.
  • Maximum heating rate: 120°C/min.
  • Supporting 32 oven heating gradients, 33 constant temperature platforms, and possible to realize programmed
    temperature rise and programmed temperature rise.
  • Maximum running time: 999.99 min.
  • Oven cooling rate: 6 minutes from 450°C to 50°C, at a room temperature of 20°C.
  • Ambient temperature sensitivity: 1°C change in ambient temperature, change in column oven average temperature <
    0.01°C.

Electronic gas circuit control :

  • Standardized pressure and temperature compensation functions.
  • Electronic pressure control with ruby damping, with pressure control accuracy at ± 0.001 psi.
  • Pressure units are available in psi, kPa, bar.
  • Programed pressure rise/flow rise: Up to forth order. Supported carrier and makeup gas types: Ar, N., He, H2. Stability of carrier gas flow rate: Less than 1%/10 min.

Injection port :

  • Supports installation of up to two injection ports.
  • Fully electronic gas circuit control, supporting pressure compensation and temperature compensation.
  • Injection port type: Ultra- deactivated split/splitless injection port.
  • Compatible with all capillary columns (inner diameter:0.1 mm- 0.53 mm).
  • Split ratio up to 12500:1 to avoid column overload.
  • Supporting, split, splitless, pulse split, pulse splitless injection and other injection modes.
  • Maximum operating temperature: 450°C.
  • Temperature setting accuracy: ≤0.1°C.
  • Carrier gas saving mode, making it possible to reduce gas consumption without compromising chromatographic
    separation effects.
  • Electronic septum purge flow control to effectively eliminate chromatographic ghost peaks.
  • Flow control range 0-500 mL/min (N), 0-1250 mL/min (H, or He).
  • Injection port supporting ultra-deactivated processing, eliminating active sites on surface and improving detection sensitivity, peak shape and reproducibility of active substances.

Detector

  • Supporting installation of up to four detectors.
  • All detectors supporting electronic gas circuit control and electronic switch control, as well as pressure compensation and
    temperature compensation

Flame lonization Detector (FID) :

  • Minimum detection limit: <1.2×10 g C/s.
  • Linear dynamic range: 10′.
  • Baseline noise: <1×10″A
  • Baseline drift: <1×10″ A within 30 min.
  • Signal acquisition frequency: 10 Hz, 20 Hz, 50 Hz, 100 Hz, and 200 Hz available.
  • Auto ignition and flameout protection.
  • Hydrogen leak protection function.
  • Maximum operating temperature: 450°C.

Electron Capture Detector (ECD) :

  • Minimum detection limit: <4.2×10 g/mL.
  • Linear range: 10″.
  • Baseline noise: <0.02 mV.
  • Baseline drift: <0.05 mV within 30 min.
  • Data acquisition rate: 10Hz, 20Hz, 50Hz, 100Hz, and 200Hz available. Radioactive source: ẞ ray from 63Ni (s10 mCi).
  • Maximum operating temperature: 350°C.

Flame Photometric Detector (FPD) :

  • Single wavelength flame photometric detector, with highselectivity and sensitivity for sulfur or phosphorous compounds.
  • Minimum detection limit: <9x10g/s (P); <5x10g/s (S) Linear range: > 10P); QB 10 (S)
  • Baseline noise: <2×10″A.
  • Baseline drift: 100 A within 30 min.
  • Maximum operating temperature: 350°C.

Auxiliary electronic flow circuit control :

  • Two positions on the rear of the GC of the GC 2000 for the auxiliary electronic flow circuit control modules.

Environmental conditions :

  • Operating ambient temperature: 5°C-35°C.
  • Operating ambient humidity: 5%-95%.
  • Limit temperatures for storage: -40°C to 70°C.
  • Power supply requirements: Voltage: AC 220 ( +10%) V; Frequency: 50 (±10%) Hz.

Safety and Regulatory certifications :

  • GB 4793.1-2007 Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General 
  • GB/T 15479-1995 Technical requirements and test methods of insulation resistance and insulating strength for use inindustrial process measurement and control instruments.

Other specifications :

  • Height: 486 mm.
  • Width: 556.4 mm.
  • Thickness: 558 mm.
  • Weight: 38.1 kg.
  • Supporting up to 6 valves.
  • Independent heating zone, 8 heating zones in total with the column oven excluded.